Spectrometers > Non-scanning Dispersive Raman Spectrometers

Description

Typical Raman signal is very weak, so that very high sensitivity spectrometers are needed to measure it. This is where P&P Optica's high performance systems fit in. Our volume phase holographic (VPH) diffraction gratings and transmission based design spectrometers are ideal for Raman applications. The non-scanning Raman spectrometer allows for the measurement of weak signals from dynamically changing samples.

Our Raman spectrometers are an extension of the wide spectral range high performance systems which were designed to deliver high throughput with minimal signal aberrations. For Raman applications, the VPH grating dispersion was increased significantly to increase the spectral resolution and to decrease the spectral range. The optics of the system were also modified to provide even better aberration control - thus allowing for the use of very tall slits and high detectors.

Figure 1. Smaple Raman spectrum obtained with P&P Optica dispersive, non-scanning spectrometer (top) and a typical Raman FT system

What does this mean to a Raman spectrometer user?

P&P Optica Raman spectrometers collect much more light than typical non-scanning Raman systems. That gives you ability to measure signals not measurable before, as the signal to noise (SNR) of the system improves greatly.

Our Raman spectrometer can collect light from slits up to 12 mm in length, providing high signal without sacrificing the resolution of the system. The low aberration optics permit the use of very thin slits, down to 10 microns and detector pixels of 12 microns - giving very high spectral resolution in a non-scanning spectrometer.

Available Systems

Product Name1 Units Excitation wavelength Shift from excitation Start of spectrum Spectral band End of spectrum Dispersion per pixel Spectral resolution2
RASPPEC-532nm-F cm-1
18797
70
18727
9398
9398
5.14
10.27
nm
532
2
534
530
1064
0.29
0.58
RASPPEC-532nm-HR cm-1
18797
53
18744
3453
15344
1.89
3.77
nm
532
2
533
118
652
0.06
0.13
RASPPEC-785nm-F cm-1
12739
73
12666
3793
8946
2.07
4.15
nm
785
5
790
328
1118
0.18
0.36
RASPPEC-785nm-HR cm-1
12739
65
12674
2147
10591
1.17
2.35
nm
785
4
789
155
944
0.09
0.17
RASPPEC-1064nm-F cm-1
9398
70
9328
3618
5780
3.53
7.07
nm
1064
8
1072
658
1730
0.64
1.28
RASPPEC-1064nm-HR cm-1
9398
53
9346
1565
7834
2.48
4.96
nm
1064
6
1070
207
1277
0.38
0.75

1 Modular design allows for custom camera and detector fitting as required by the customer.

2 Performance data for excitation bandwidth <= 1cm-1

System Options

These are examples of Raman system configurations offered by P&P Optica. Please contact us for more information regarding these and any other systems.

High Performance Raman Spectrometer for 785 nm excitation wavelength (RASPPEC-785nm-F)

  • High performance spectrometer with VPH grating
  • <0.01% scattered light level
  • 300cm-1 spectral range
  • 3.25cm-1/pixel mean dispersion
  • installed and tested 500mW, 785nm, 3.5cm-1 line width LD
  • IR enhanced Hamamatsu linear array detector, 1TEC cooled, SNR>1000
  • 1.5m, 300-3000cm-1, 0.2mm core diameter fiber optic probe

Option 1: Back thinned Hamamatsu detector, 1TEC cooled, 1.6mm, 2D array, SNR>3000

Option 2: Back thinned Hamamatsu detector, 2TEC cooled, 1.6mm, 2D array, SNR>3000 reduced dark noise for weak signals

Option 3: 785 nm 300mW laser, 1.0cm-1 line width

Option 4: 786 nm 600mW laser, 1.0cm-1 line width

Option 5: 5m, 200-3000cm-1 fiber optic probe

Option 6: < 50-3000cm-1 Raman probe with fiber optic delivery and light redistribution, matched to a spectrometer with back thinned Hamamatsu detector, 2TEC cooled, 6mm tall, 2D array, SNR>10,000 and reduced dark noise for weak signals

Option 7: Sample holder for variable diameter cylindrical test tubes

High Resolution, High Performance Raman Spectrometer for 785 nm excitation wavelength (RASPPEC-785nm-HR)

  • High performance spectrometer with VPH grating
  • <0.01% scattered light level
  • 2000cm-1 spectral range
  • 2cm-1/pixel mean dispersion
  • installed and tested 300mW, 785nm, 1cm-1 line width LD
  • Back thinned Hamamatsu detector, 2TEC cooled, 6mm tall, 2D array, SNR>10,000 reduced dark noise for weak signals

Option 1: < 50-2000cm-1 Raman probe with fiber optic delivery and light redistribution, matched to a spectrometer with back thinned Hamamatsu detector, 2TEC cooled, 6mm tall, 2D array

Option 2: Sample holder for variable diameter cylindrical test tubes

Option 3: installed and tested 600mW, 785nm, 1cm-1 line width LD

Technical Specifications

Parameter Value
size 32cm L x 12cm W x 10 cm H
weight typically <2 kg without detector
F/number F/3
Spectral range (free of second order effects) 30% of an octave in the 350 nm to 2500 nm range
Spot diameter 19 µm @ 1000 nm in 25.4 x 9 mm2 area
Spectral resolution (0.024 mm wide pixels) 0.13 nm @ 532 nm and 0.17 785 @ 2000 nm
Linear dispersion 30% of an octave per inch
Maximum detector size 25.4 x 12.7 mm
Dynamic range of a single 0.024 by 0.024 mm2 pixel > 3,000
Dynamic range, binning along 2 mm slit > 55,000 (detector dependent)
Dynamic range, binning along 8 mm slit > 300,000 (up to 1,000,000 with averaging over a number of measurements)